3.13 Connecting MSB devices
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
855 Robotic Titrosampler
3.13.1 Connecting dosing devices
Three dosing devices can be connected to the instrument.
The types of dosing devices that are supported are:
■ 800 Dosino
■ 700 Dosino
■ 805 Dosimat
■ 685 Dosimat
Warning
If a Dosino is connected to the 855 Robotic Titrosampler then the con-
nection cable must be equipped with a T.2400.102 ferrite core. The ferrite
core reduces any interference voltages that may occur and thus ensures
compliance with strict EMC standards pursuant to applicable technical
norms, see Chapter "Technical Data".
Proceed as follows:
1
Mounting ferrite core
Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.
2
Connect a dosing device
■ Exit the control software.
■ Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.
■ Start the control software.
USB 1
Contr.
MSB 2
MSB 3
T.2400.102
Figure 23 Connecting a dosing device
Comments to this Manuals