3 Operating tutorial
792 Basic IC
46
3.2 Preparations
Before you start this brief tutorial, the entire IC system must be correctly
installed as described in
section 2.
In what follows, the most important
points for the installation are described once again
(for details, see the
sections mentioned).
1 Install 792 Basic IC
Setting up instrument
section 2.2
Installing and connecting detector block
section 2.3.1
Mounting syringe and aspirating tubing
section 2.3.2
Mains connection
section 2.4
Connection to PC
section 2.5
2 Prepare eluent
Preparing eluent:
2.4 mmol/L NaHCO
3
/ 2.5 mmol/L Na
2
CO
3
/
2 % Acetone in dist. H
2
O
Microfiltering and degassing eluent
section 5.1.3
3 Install high-pressure pump
Removing transport security screws
section 2.6.1
Mounting pulsation dampener
section 2.6.2
Installing eluent supply
section 2.6.3
Degassing pump
section 2.6.4
4 Connecting separating column
Connecting IC cation column
section 2.7.7
Conditioning system
section 2.9.1