5.2 Maintenance and servicing
792 Basic IC
149
6 Mounting the pump head
x
Reassemble the components of the pump head
34
3434
34
as shown
in
Fig. 16
. Tighten the screw
81
8181
81
by hand. First screw in piston
cartridge
85
8585
85
manually until the stop is reached and then use
a wrench to turn it through a further 15°. Firmly tighten the two
valve screw holders
92
9292
92
with a wrench.
x
Reattach pump head
34
3434
34
to the pump with the help of the 4
fixing screws
33
3333
33
. Firmly tighten them with the 6.2621.030 Al-
len key.
To ensure that the pump head is not positioned wrongly, the holes at
the rear for the clamping bolts have different depths, i.e. 1 clamping
bolt is longer than the rest. The deepest hole must naturally accom-
modate the longest bolt. If this is not the case, the pump will not
function properly.
x
Screw connection capillary
35
3535
35
to pump head
34
3434
34
(see
Fig. 3
).
x
Connect aspirating tubing to aspirating capillary
32
3232
32
at the
pump head
34
3434
34
.
Inlet valve Outlet valve
6.2824.090 6.2824.080
95
9595
95
96
9696
96
97
9797
97
98
9898
98
99
9999
99
100
100100
100
101
101101
101
95
9595
95
96
9696
96
97
9797
97
100
100100
100
99
9999
99
98
9898
98
101
101101
101
Fig. 18
: Components of inlet valve
90 and
outlet valve
91
95
9595
95 Valve housing
96
9696
96 Sealing ring
(black)
97
9797
97
Sleeve
98
9898
98
Sapphire sleeve
The bright side must point towards
the sapphire sphere
99
9999
99
Sapphire sphere
100
100100
100
Ceramic holder for sapphire
sphere
The large recess must point in the
direction of the sapphire sphere
101
101101
101
Seal
The larger opening must point
outwards