NIRS XDS MasterLab Analyzer Manual 8.921.8003EN / 2013-12-19
8 ▪▪▪▪▪▪▪ 2 Site Readiness Like most precision instruments, the MasterLab Analyzer (RCA) is sensitive to environmental conditions that can affect it
98 ▪▪▪▪▪▪▪ When all standards are scanned, results are posted as shown. When Vision has completed the test, the tabulated results may be printed. Th
▪▪▪▪▪▪▪ 99 The four standards are plotted out in the USP format, showing repeatable response for each standard. The x-axis, A REF, is from the initia
100 ▪▪▪▪▪▪▪ The view above shows Transmission gain on a typical MasterLab Analyzer. The gain program measures amplifier gain factor and voltage thro
▪▪▪▪▪▪▪ 101 From the Diagnostics menu bar, select Low Flux Test. Vision asks if the user wishes to use an external sample (standard) for the test.
102 ▪▪▪▪▪▪▪ 8 Instrument Diagnostics, Reflectance Vision provides diagnostics for instrument setup, which must be performed before use of the instru
▪▪▪▪▪▪▪ 103 Use Window Correction (XDS only): This feature is used only with XDS Process instruments that operate with Transmission Pair probes. Do n
104 ▪▪▪▪▪▪▪ minutes, depending upon whether the instrument has been off, or if the transmission detector was used previously. Once Vision shows the i
▪▪▪▪▪▪▪ 105 The results screen shown above is typical. Peak positions for the reference materials are located using a peak-finding algorithm. These
106 ▪▪▪▪▪▪▪ Reference Standardization. The benefit of this system is that only one calibration (or library) is required, and serves both sampling lev
▪▪▪▪▪▪▪ 107 When Vision first communicates with the instrument (or after disconnection and re-connection) this screen is shown. Vision must establish
▪▪▪▪▪▪▪ 9 standard RJ-45 type cable, such as CDW #074092, available from CDW Computer Centers, Inc. The instrument detects network capability and opt
108 ▪▪▪▪▪▪▪ Vision plots a spectrum of the ceramic instrument reference, as shown. (The spectrum will be light blue on the screen – it is plotted in
▪▪▪▪▪▪▪ 109 The correction spectrum represents the amount of spectral correction required to provide a virtual 100% reflectance reference at each dat
110 ▪▪▪▪▪▪▪ 8.1.3 Reference Standardization, Reflectance, “Use Iris” Position Next we will perform Reference Standardization in the “Use Iris” mode.
▪▪▪▪▪▪▪ 111 If Vision displays the Instrument Configuration screen again, click “OK to accept the instrument identification. Vision briefly displays
112 ▪▪▪▪▪▪▪ Vision performs the same operations as done on the previous Reference Standardization. Click “OK”. The Certified 80% Reflectance Standa
▪▪▪▪▪▪▪ 113 A final spectrum (green when plotted on-screen) is plotted to verify that the corrected spectrum produces the same results as the Certifi
114 ▪▪▪▪▪▪▪ 8.1.4 Instrument Calibration, Reflectance Instrument Calibration uses a NIST-traceable, stable, standard, of known wavelength response,
▪▪▪▪▪▪▪ 115 Vision requests the “Standard File” for the Certified Wavelength Standard. Place the mini-CD into the CD drive, and select the standard
116 ▪▪▪▪▪▪▪ Information about the reflectance wavelength standard: The wavelength standard used is directly traceable to NIST SRM-1920a, through dire
▪▪▪▪▪▪▪ 117 8.1.5 IPV Setup, Reflectance (Instrument Performance Verification) IPV Setup is provided as a method to record initial instrument respon
10 ▪▪▪▪▪▪▪ 3 XDS Instrument Connection The XDS instrument may be connected to the host computer in one of two ways: If the XDS Instrument will be us
118 ▪▪▪▪▪▪▪ Vision requests a “Standard File”. This is provided on a mini-CD, packed in the wooden box with the standards. Insert this mini-CD into t
▪▪▪▪▪▪▪ 119 Place the 99% standard into the tray. The label edge should be parallel to the long edge of the tray. This enhances repeatability. Once
120 ▪▪▪▪▪▪▪ 8.2 Evaluation Diagnostics, Reflectance Evaluation Diagnostics are used to verify that the instrument is operating within allowable para
▪▪▪▪▪▪▪ 121 Because only the right two quadrants of the screen will be updated, many users “pull the screen over” to see better, as shown on the next
122 ▪▪▪▪▪▪▪ Noise Summary displayed using the OpQual tab. The OpQual tab brings up the display shown. This shows results of the Noise Test for each
▪▪▪▪▪▪▪ 123 8.2.2 Wavelength Certification, Reflectance Wavelength Certification is used to confirm the peak positions of the instrument to a define
124 ▪▪▪▪▪▪▪ Double-click the lower right quadrant to see the full report. The first tab shows the Wavelength positions, as found. They are compare
▪▪▪▪▪▪▪ 125 The wavelengths used for the instrument wavelength profile are well-defined, stable peaks in the wavelength standard. These are the same
126 ▪▪▪▪▪▪▪ The wavelength peaks used here are not the NIST peak positions, though they are within +/- 1.0nm of the NIST peaks. The peaks used in In
▪▪▪▪▪▪▪ 127 provide an ongoing record of performance. The test uses the same standards used in IPV Setup. Photometric Test compares current spectra o
▪▪▪▪▪▪▪ 11 • The XDS instrument appears just like a network printer (or other peripheral device) on the LAN system. It generates no signals, and onl
128 ▪▪▪▪▪▪▪ Vision displays the wavelength regions for test. For reflectance instruments, these regions give a good overall picture of instrument per
▪▪▪▪▪▪▪ 129 Insert the 99% standard into the tray as shown, with the label parallel to the long axis of the tray. When the 99% standard has been sca
130 ▪▪▪▪▪▪▪ The five standards are plotted out in the USP format, showing repeatable response for each standard. The x-axis, A REF, is from the initi
▪▪▪▪▪▪▪ 131 To start Gain Adjust, click on Diagnostics, then Gain Adjust. The instrument is connected and in communication for this to function. Th
132 ▪▪▪▪▪▪▪ From the Diagnostics menu bar, select Low Flux Test. Vision asks if the user wishes to use an external sample (standard) for the test.
▪▪▪▪▪▪▪ 133 Click on the tab marked “Summary” to see the summarized results as compared to acceptance specifications. Vision reports a pass or fail
134 ▪▪▪▪▪▪▪ 9 Instrument Maintenance Instrument maintenance is quite simple on the XDS MasterLab Analyzer. The optical enclosure is sealed to preven
▪▪▪▪▪▪▪ 135 9.2 Fan Filter Replacement The fan filter should be inspected at least monthly. (If installed in a dusty or dirty environment, it should
136 ▪▪▪▪▪▪▪ Using the new filter material, gently insert it into the filter cover. It should fit easily in any direction, as it is symmetrical. Be s
▪▪▪▪▪▪▪ 137 Turn off power at the power switch located on the back panel. Unplug the power cord. While this is not essential, it makes access easier
12 ▪▪▪▪▪▪▪ 3. 3. Upon being powered up on a LAN, the XDS instrument requests a “dynamic” IP address from the DHCP server which controls the LAN. Thi
138 ▪▪▪▪▪▪▪ This orientation photo shows the area inside the back cover. The lamp box is at the lower left. The fan and heat sinks, shown to the righ
▪▪▪▪▪▪▪ 139 Gently grasp the corners of the lamp mounting plate. Push in about 1mm, then rotate clock-wise (to the right) about 5 mm. The lamp mounti
140 ▪▪▪▪▪▪▪ Position the lamp with the black arrow up as shown. Gently insert the lamp and mounting plate into the lamp housing as shown. Engage the
▪▪▪▪▪▪▪ 141 Carefully re-install the back cover, lining up the captive screws with the mating threaded holes. Tighten gently by hand. Do not over-ti
142 ▪▪▪▪▪▪▪ Perform tests as specified in the prior sections of this manual. These normally include the following: • Wavelength Linearization • Per
▪▪▪▪▪▪▪ 143 Remove the power cord for access to the fuse. Use a tool (such as an Allen wrench) to pull the housing open. There is a small slot, und
144 ▪▪▪▪▪▪▪ When finished, re-install the power cord. Turn on AC power and re-establish instrument communication. 9.5 Maintenance Log Vision provid
▪▪▪▪▪▪▪ 145 This screen shows entries, sorted by record number. To re-sort by another parameter, click on the column heading. 9.6 Window Cleaning T
146 ▪▪▪▪▪▪▪ After cleaning, allow any residual alcohol to evaporate off. Then use the sample brush from the accessory kit to gently brush away any li
▪▪▪▪▪▪▪ 147 10 Validation Tools Validation is an overriding concern in the pharmaceutical marketplace. In the United States, manufacturers must foll
▪▪▪▪▪▪▪ 13 3.4.1 Microsoft Windows Firewalls The Microsoft Windows® Firewall on the PC may interfere with Vision communication. To assure communicat
148 ▪▪▪▪▪▪▪ Wavelength Accuracy, Photometric Certification, and Wavelength Certification. These factory tests are performed under controlled, ideal c
▪▪▪▪▪▪▪ 149 Independent assessment, not performed by instrument user • Meets independent test requirements of CFR 21 and most auditing methods. Ful
150 ▪▪▪▪▪▪▪ the pharmaceutical industry. Documents were selected (from the enormous quantity of records in our logs) based upon the patterns set duri
▪▪▪▪▪▪▪ 151 This screen shows many of the setup items to be set by the System Manager for 21 CFR Part 11 Compliance. To access this screen in Vision,
152 ▪▪▪▪▪▪▪ 11 Safety and Electrical Certification The XDS MasterLab Analyzer and all associated components have been tested for CE (Communite Europ
▪▪▪▪▪▪▪ 153 12 Troubleshooting The XDS MasterLab Analyzer is a dependable, trouble-free instrument, designed for many years of service in your labor
154 ▪▪▪▪▪▪▪ 2 No communication between Vision and the instrument. 1. Verify that the RJ-45 cable is plugged in at both the instrument and at the net
▪▪▪▪▪▪▪ 155 3 MasterLab Analyzer appears to “lock up” upon successful connection. 1. Upon successful connection, the instrument “lost track” of whic
156 ▪▪▪▪▪▪▪ 5 Instrument fails Wavelength Linearization 1. Instrument Reference is not in correct position. Verify that reference paddle is visible
▪▪▪▪▪▪▪ 157 7 Instrument fails Instrument Calibration. 1. Verify that the Wavelength Standard is properly positioned in the holder when prompted by
14 ▪▪▪▪▪▪▪ Internet Protocol (IP) and it is the first version of the protocol to be widely deployed. Together with IPv6, it is at the core of standar
158 ▪▪▪▪▪▪▪ 9 Instrument cooling fans are operating at maximum rate. 1. Verify that air intake on side is not blocked by other equipment. Leave at l
▪▪▪▪▪▪▪ 159 12 X-Y Sampling Mechanism crashes into Iris Adapter, causing Limit Switch Error Remove Iris Adapter. Click “Use Iris” in Vial Dialog box.
160 ▪▪▪▪▪▪▪ 16 Vision reports “Auto Linearization failed” Auto Linearization is a program to maintain correct wavelength position of the instrument.
▪▪▪▪▪▪▪ 161 13 Lifting and transporting 1 Always separate the module from the monochromator (shown) before lifting or transporting the instrument. W
162 ▪▪▪▪▪▪▪ 13.1 Instructions for MasterLab® Shipping Position When shipping the MasterLab Analyzer, it is vital to set the sample carriage to the “
▪▪▪▪▪▪▪ 163 When re-entering the project to run the instrument, be sure to select the correct “Use Tray” or “Use Iris” setting in Vision. This assure
164 ▪▪▪▪▪▪▪ 14 Index Accessory Kit 21 Aperture 26, 27, 76 Assembly 21, 46 Auto-Linearization 50, 68, 102, 155, 160 Bias 89, 122 Boat 28, 50, 6
▪▪▪▪▪▪▪ 165 Vibration 8 Voltage 160 Wavelength Certification 69, 83, 90, 93, 94, 103, 116, 118, 123, 124, 125, 148 Wavelength Linearization 74, 7
▪▪▪▪▪▪▪ 15 2. Highlight “NIRSystems XDS-series Instrument Driver” as shown, then click on “Configure”. Information: At this point, Vision requests a
16 ▪▪▪▪▪▪▪ 1. Vision cannot see any instrument on the connection path. Click on the “down arrow” at the right side of the empty field to see if inst
▪▪▪▪▪▪▪ 17 • Click on Start, then Control Panel • Double-click on Network Connections • Double-click on Local Area Connection • Click on Local Ar
18 ▪▪▪▪▪▪▪ Network Solution 2: Returning to the Local Area Connection Status dialog box, note these items for the computer: • Address Type: (should
▪▪▪▪▪▪▪ 19 Network Solution 4: Verify that the Firewall on the computer has Vision loaded as an “exception”. This is located under “Control Panel”,
20 ▪▪▪▪▪▪▪ 3.6.2 Direct Connection Troubleshooting Overview If no XDS instrument shows as “available”, the computer may need to be configured for t
▪▪▪▪▪▪▪ 21 4 Assembly of the Instrument The XDS Analyzer will be assembled and installed by a trained representative of Metrohm. This person will pe
22 ▪▪▪▪▪▪▪ 1 Load Vision Spectral Analysis Software onto the computer designated to operate the XDS instrument. 2 Place the monochromator on the lab
▪▪▪▪▪▪▪ 23 5 Insert the AC power cable into the AC power block as shown. 6 Attach the RJ-45 cable to the network connector on the instrument. If usi
24 ▪▪▪▪▪▪▪ 9 Lift the release handle on the monochromator and engage the module “catches” to the locking togs on the monochromator. (Module not shown
▪▪▪▪▪▪▪ 25 13 When all the above assembly is finished, turn on the power switch on the monochromator. It is located on the lower surface, on the righ
26 ▪▪▪▪▪▪▪ 5 MasterLab Analyzer Overview The XDS MasterLab Analyzer is designed for several types of versatile analysis, from tablet sampling, to vi
▪▪▪▪▪▪▪ 27 Large Sample Cell, Reflectance • Detector Selection: Reflectance • Wavelength Range: 400-2500 nm • Preferred Spot Size: 17.25 mm • Ape
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 [email protected] www.metrohm.com NIRS XDS MasterLab Analyz
28 ▪▪▪▪▪▪▪ Transmission Reflectance Use Tray Use for multiple tablets, scanning in transmission. The default spot size (9.5mm) is appropriate for
▪▪▪▪▪▪▪ 29 5.1 Introduction to Tablet Analysis The MasterLab Analyzer is designed to provide sampling for trays of tablets, like that shown at right
30 ▪▪▪▪▪▪▪ that in tablet transmission, some sample positions are not available because the reference standard takes up one or two tablet positions.
▪▪▪▪▪▪▪ 31 5.2 Creation of a Custom Vial Size Most users will work with one of the standard vial sizes (or tablet layouts) offered. However, for set
32 ▪▪▪▪▪▪▪ A “Tray ID” must be assigned. The normal convention is to use the diameter, then the number of vials in a row, and a column. You may use y
▪▪▪▪▪▪▪ 33 These dimensions are the center of each vial opening in the tray. Vision uses these coordinates to position each vial during sampling. Not
34 ▪▪▪▪▪▪▪ The adjustment ranges from 9.5mm to 17.25mm, in adjustment steps of 0.25mm. The effective scanned sample areas at various spot sizes are
▪▪▪▪▪▪▪ 35 When using the default spot size of 17.25mm, the sample diameter should be at least 21.6mm (0.85 inch) as represented by the gray circle i
36 ▪▪▪▪▪▪▪ To insert the sample tray into the instrument, note the round guides on the sample carrier. They are shown in this photo – the pencil poin
▪▪▪▪▪▪▪ 37 1. Remove the black covers from the adjustment post on the instrument, and the adjustment screw packed with the standards tray. 2. Thr
Teachware Metrohm AG CH-9100 Herisau [email protected] This documentation is protected by copyright. All rights reserved. Although all
38 ▪▪▪▪▪▪▪ 5. The vertical post of the adjustment is marked in 1mm (about 0.040”) increments. The rotary knob is marked in 50 increments. One markin
▪▪▪▪▪▪▪ 39 Once set, the height will remain constant, even when the transmission detector is moved to the storage position as shown. When rotated bac
40 ▪▪▪▪▪▪▪ 5.6 Universal Tablet Tray Part Number 3013-243-01 The XDS MasterLab® Analyzer is shipped with one “Universal Tablet Tray” which may be us
▪▪▪▪▪▪▪ 41 of the Universal Tablet Tray. (Remember that spectral intensity changes somewhat with detector height.) Bore sizes (diameter), for tablet
42 ▪▪▪▪▪▪▪ 5.7 Tablet Tray Unloader Because tablets fit snugly in the tray, and may be difficult to remove, a device has been designed to make unloa
▪▪▪▪▪▪▪ 43 The vacuum lifting device is easily controlled by blocking or releasing the hole on the wand. Cover the hole to pick up the tablet. To rel
44 ▪▪▪▪▪▪▪ When using the Iris Adapter, Vision provides a command to move the sample carriage to a neutral position, as shown at right. First, remove
▪▪▪▪▪▪▪ 45 When ready, and all interfering items are removed, click “OK” in this dialog box. When scanning stationary objects, the “Use Iris” select
46 ▪▪▪▪▪▪▪ Use the iris to center samples in the window area. Push the handle to the right to center the sample, then to the left to move the iris fr
▪▪▪▪▪▪▪ 47 6 Vision Software: Connection to the Instrument This section describes communication between the computer (with Vision Software loaded) a
▪▪▪▪▪▪▪ 3 Table of contents 1 Introduction to XDS™ MasterLab Analyzer ... 5 2 Site
48 ▪▪▪▪▪▪▪ A new project must be created. The “project” is used to store data and calibrations for a given type of analysis. Multiple projects may b
▪▪▪▪▪▪▪ 49 This box directs the user to select the instrument IP (Internet Protocol) address. Use the drop-down arrow of the IP Address box, and sele
50 ▪▪▪▪▪▪▪ Verify that these selections are clicked: Instrument must stabilize before data acquisition Use Auto-Linearization Reference Standardizati
▪▪▪▪▪▪▪ 51 If the user clicks on “Modify/ View” from the Select Tray Dialog box, this screen is shown. If “New” is clicked, all fields are blank, for
52 ▪▪▪▪▪▪▪ The instrument “self-identifies” as an XDS system with the Tablet Module. The user should enter a logical name for the DCM. (Spaces are OK
▪▪▪▪▪▪▪ 53 Next, select this DCM from the selection box. Click “OK” to connect to the instrument. Once connected, verify that the amber “Communicatio
54 ▪▪▪▪▪▪▪ If a Reference Standardization does not exist for the selected configuration, Vision will display this message. The user should perform b
▪▪▪▪▪▪▪ 55 instrument operation. Please proceed to the next step. In tablet transmission, Vision stores yet another Reference Standardization, for th
56 ▪▪▪▪▪▪▪ Note that the Vial Dialog box is also called when either of these actions is performed: 1. The user clicks on the “Sample” Icon, or 2.
▪▪▪▪▪▪▪ 57 Also note that 20 vials are shown, in a 4 x 5 configuration. This came from the Select Tray Dialog box. (It is shown at right, as a remin
4 ▪▪▪▪▪▪▪ 7.1.2 Reference Standardization, Tablet Transmission ... 76 7.1.3 Instrument Calibration
58 ▪▪▪▪▪▪▪ Entry Method 2: Double-click on the upper part of the vial position, which brings up the box shown: The Sample ID may be typed into this
▪▪▪▪▪▪▪ 59 The cell entry method is shown above, highlighted by the red border. The user may type a Sample ID, then press “Enter” to move down to th
60 ▪▪▪▪▪▪▪ Highlight the Sample ID fields to be pasted as shown above. Click on the “Paste” button in the dialog box. (It is outlined in red below.)
▪▪▪▪▪▪▪ 61 There is a field for comments beside each sample. To see this field, click the small arrow at the bottom right, and the field will be visi
62 ▪▪▪▪▪▪▪ To delete multiple Sample ID names, highlight those fields, and use the “Delete” key on the PC keyboard. Vision will ask for confirmation,
▪▪▪▪▪▪▪ 63 below, this is automatic. At the bottom of the Vial Dialog screen is a checkbox marked “Use Order”. If this box is checked, the MasterLa
64 ▪▪▪▪▪▪▪ In the example above, the order is out of sequence for samples, 7, 9, and 16. If “Use Order” is left unchecked, Vision proceeds in order
▪▪▪▪▪▪▪ 65 This setting defines the number of sample spectra to be taken per reference scan. With multiple samples, it is cumbersome to take a refere
66 ▪▪▪▪▪▪▪ Use Iris: If the user needs to scan a sample using the Iris Adapter, this box should be clicked. (This resets the choice made in the “Sel
▪▪▪▪▪▪▪ 67 If the user needs to scan a sample using the large Sample Cell (boat), this box should be clicked. (This resets the choice made in the “Se
▪▪▪▪▪▪▪ 5 1 Introduction to XDS™ MasterLab Analyzer Thank you for selecting the XDS MasterLab Analyzer, manufactured by FOSS. This instrument is the
68 ▪▪▪▪▪▪▪ 7 Instrument Diagnostics, Tablet Transmission Vision provides diagnostics for instrument setup. These diagnostics must be performed befor
▪▪▪▪▪▪▪ 69 This is a method to adjust the instrument wavelength profile to an external, traceable wavelength standard. This feature must be used to a
70 ▪▪▪▪▪▪▪ A new DCM screen will be shown. The “Method” name will be blank, and the “Detector” may be marked as “Reflectance”. Proceed with these st
▪▪▪▪▪▪▪ 71 To verify this connection, the user may click on Diagnostics, Instrument Configuration. Note that the Amplifier Type is listed as “Transm
72 ▪▪▪▪▪▪▪ It is easier to install the aperture when the sample carriage is in the “Iris Position”. The unit may be in this position from previous w
▪▪▪▪▪▪▪ 73 To re-install the tablet transmission reference, mount it onto the rail as shown, and screw down the captive thumbscrew about 4-5 turns.
74 ▪▪▪▪▪▪▪ The directions that follow include spectroscopic calibrations, which are used to “normalize” the instrument, to minimize differences betwe
▪▪▪▪▪▪▪ 75 These peak positions are not meant to be traceable, as the true wavelength calibration of the instrument is done on an external standard,
76 ▪▪▪▪▪▪▪ If Wavelength Linearization fails for any reason, check these items: • Tablet Transmission Detector in correct position • Tablet Transmi
▪▪▪▪▪▪▪ 77 When Vision first communicates with the instrument (or after disconnection and re-connection) this screen is shown. Vision must establish
6 ▪▪▪▪▪▪▪ the sections on Reference Standardization. This instrument uses near-infrared (NIR) spectral energy to illuminate the sample. By measuring
78 ▪▪▪▪▪▪▪ Vision plots the measured difference between the instrument tablet transmission reference and the certified standard in the Transmission S
▪▪▪▪▪▪▪ 79 Vision plots the spectrum of the final corrected reference spectrum in dark green. This spectrum should appear exactly the same as the cal
80 ▪▪▪▪▪▪▪ From the menu bar, click on Diagnostics, then Instrument Calibration. Vision requests that the aperture and Tablet Transmission Standard
▪▪▪▪▪▪▪ 81 Vision requests the “Standard File” for the Certified Wavelength Standard. Place the mini-CD into the CD drive, and select the standard f
82 ▪▪▪▪▪▪▪ At the end of the test this dialog box is displayed. Print confirmation of test results if needed for a log book or other file. Click “Pri
▪▪▪▪▪▪▪ 83 to illustrate that the wavelength peaks of SRM-2035 remain constant, even when neutral density filters are added to increase absorbance le
84 ▪▪▪▪▪▪▪ under Configure, Options. That is, if the IPV Setup file is acquired with Reference Standardization switched on, and with Window Correctio
▪▪▪▪▪▪▪ 85 Vision verifies the tray type – select the Tablet Transmission Standard, as highlighted. Verify “Use Tray”. Click on “Select”, then “Close
86 ▪▪▪▪▪▪▪ Vision takes an instrument reference. The reference spectrum is shown in red. Note that Reference Standardization is not shown as applied
▪▪▪▪▪▪▪ 87 7.2 Evaluation Diagnostics, Tablet Transmission Evaluation Diagnostics are used to verify that the instrument is operating within allowab
▪▪▪▪▪▪▪ 7 The instrument enclosure is completely sealed to prevent contamination by dust or other substances. The cooling fans operate outside the ma
88 ▪▪▪▪▪▪▪ Because only the right two quadrants of the screen will be updated, many users “pull the screen over” to see better, as shown on the next
▪▪▪▪▪▪▪ 89 When Performance Test is finished, a message box is displayed to indicate test completion and status. All measured values are compared wit
90 ▪▪▪▪▪▪▪ protected and are moved by software command, transparent to the user. When Performance Test is run, the relative wavelength positions and
▪▪▪▪▪▪▪ 91 Vision asks the user to select the Standard File. This file is on the mini-CD, as shown. This file may be copied into the C:\Vision direc
92 ▪▪▪▪▪▪▪ Double-click the lower right quadrant to see the full report. The first tab shows the Wavelength position, as found. Only one is used w
▪▪▪▪▪▪▪ 93 Note that the peak position is shown only to two digits after the decimal point. Any further digits are not considered significant, and ar
94 ▪▪▪▪▪▪▪ In some cases the user may see the message shown. This indicates that the instrument has passed Wavelength Certification on the NIST-defin
▪▪▪▪▪▪▪ 95 Vision requests a “Standard File”. For Photometric Test, use the TSSVxxxx.da file stored in the Vision directory. This file was created du
96 ▪▪▪▪▪▪▪ Vision displays the wavelength regions for test. For tablet transmission, these regions give a good overall picture of instrument performa
▪▪▪▪▪▪▪ 97 Vision verifies the tray layout as shown. Click on “OK”. Vision acquires a reference scan. Next, the T02 is scanned. It is reported at ri
Comments to this Manuals