Metrohm NIRS XDS MasterLab Analyzer User Manual

Browse online or download User Manual for Equipment Metrohm NIRS XDS MasterLab Analyzer. Metrohm NIRS XDS MasterLab Analyzer User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - XDS MasterLab Analyzer

NIRS XDS MasterLab Analyzer Manual 8.921.8003EN / 2013-12-19

Page 2

8 ▪▪▪▪▪▪▪ 2 Site Readiness Like most precision instruments, the MasterLab Analyzer (RCA) is sensitive to environmental conditions that can affect it

Page 3

98 ▪▪▪▪▪▪▪ When all standards are scanned, results are posted as shown. When Vision has completed the test, the tabulated results may be printed. Th

Page 4

▪▪▪▪▪▪▪ 99 The four standards are plotted out in the USP format, showing repeatable response for each standard. The x-axis, A REF, is from the initia

Page 5 - Table of contents

100 ▪▪▪▪▪▪▪ The view above shows Transmission gain on a typical MasterLab Analyzer. The gain program measures amplifier gain factor and voltage thro

Page 6

▪▪▪▪▪▪▪ 101 From the Diagnostics menu bar, select Low Flux Test. Vision asks if the user wishes to use an external sample (standard) for the test.

Page 7

102 ▪▪▪▪▪▪▪ 8 Instrument Diagnostics, Reflectance Vision provides diagnostics for instrument setup, which must be performed before use of the instru

Page 8

▪▪▪▪▪▪▪ 103 Use Window Correction (XDS only): This feature is used only with XDS Process instruments that operate with Transmission Pair probes. Do n

Page 9

104 ▪▪▪▪▪▪▪ minutes, depending upon whether the instrument has been off, or if the transmission detector was used previously. Once Vision shows the i

Page 10 - 2 Site Readiness

▪▪▪▪▪▪▪ 105 The results screen shown above is typical. Peak positions for the reference materials are located using a peak-finding algorithm. These

Page 11

106 ▪▪▪▪▪▪▪ Reference Standardization. The benefit of this system is that only one calibration (or library) is required, and serves both sampling lev

Page 12 - 3 XDS Instrument Connection

▪▪▪▪▪▪▪ 107 When Vision first communicates with the instrument (or after disconnection and re-connection) this screen is shown. Vision must establish

Page 13

▪▪▪▪▪▪▪ 9 standard RJ-45 type cable, such as CDW #074092, available from CDW Computer Centers, Inc. The instrument detects network capability and opt

Page 14

108 ▪▪▪▪▪▪▪ Vision plots a spectrum of the ceramic instrument reference, as shown. (The spectrum will be light blue on the screen – it is plotted in

Page 15

▪▪▪▪▪▪▪ 109 The correction spectrum represents the amount of spectral correction required to provide a virtual 100% reflectance reference at each dat

Page 16 - 3.5 Connection in Vision

110 ▪▪▪▪▪▪▪ 8.1.3 Reference Standardization, Reflectance, “Use Iris” Position Next we will perform Reference Standardization in the “Use Iris” mode.

Page 17

▪▪▪▪▪▪▪ 111 If Vision displays the Instrument Configuration screen again, click “OK to accept the instrument identification. Vision briefly displays

Page 18

112 ▪▪▪▪▪▪▪ Vision performs the same operations as done on the previous Reference Standardization. Click “OK”. The Certified 80% Reflectance Standa

Page 19

▪▪▪▪▪▪▪ 113 A final spectrum (green when plotted on-screen) is plotted to verify that the corrected spectrum produces the same results as the Certifi

Page 20

114 ▪▪▪▪▪▪▪ 8.1.4 Instrument Calibration, Reflectance Instrument Calibration uses a NIST-traceable, stable, standard, of known wavelength response,

Page 21

▪▪▪▪▪▪▪ 115 Vision requests the “Standard File” for the Certified Wavelength Standard. Place the mini-CD into the CD drive, and select the standard

Page 22

116 ▪▪▪▪▪▪▪ Information about the reflectance wavelength standard: The wavelength standard used is directly traceable to NIST SRM-1920a, through dire

Page 23

▪▪▪▪▪▪▪ 117 8.1.5 IPV Setup, Reflectance (Instrument Performance Verification) IPV Setup is provided as a method to record initial instrument respon

Page 24

10 ▪▪▪▪▪▪▪ 3 XDS Instrument Connection The XDS instrument may be connected to the host computer in one of two ways: If the XDS Instrument will be us

Page 25

118 ▪▪▪▪▪▪▪ Vision requests a “Standard File”. This is provided on a mini-CD, packed in the wooden box with the standards. Insert this mini-CD into t

Page 26

▪▪▪▪▪▪▪ 119 Place the 99% standard into the tray. The label edge should be parallel to the long edge of the tray. This enhances repeatability. Once

Page 27

120 ▪▪▪▪▪▪▪ 8.2 Evaluation Diagnostics, Reflectance Evaluation Diagnostics are used to verify that the instrument is operating within allowable para

Page 28

▪▪▪▪▪▪▪ 121 Because only the right two quadrants of the screen will be updated, many users “pull the screen over” to see better, as shown on the next

Page 29

122 ▪▪▪▪▪▪▪ Noise Summary displayed using the OpQual tab. The OpQual tab brings up the display shown. This shows results of the Noise Test for each

Page 30

▪▪▪▪▪▪▪ 123 8.2.2 Wavelength Certification, Reflectance Wavelength Certification is used to confirm the peak positions of the instrument to a define

Page 31

124 ▪▪▪▪▪▪▪ Double-click the lower right quadrant to see the full report. The first tab shows the Wavelength positions, as found. They are compare

Page 32

▪▪▪▪▪▪▪ 125 The wavelengths used for the instrument wavelength profile are well-defined, stable peaks in the wavelength standard. These are the same

Page 33

126 ▪▪▪▪▪▪▪ The wavelength peaks used here are not the NIST peak positions, though they are within +/- 1.0nm of the NIST peaks. The peaks used in In

Page 34

▪▪▪▪▪▪▪ 127 provide an ongoing record of performance. The test uses the same standards used in IPV Setup. Photometric Test compares current spectra o

Page 35 - 5.3 Spot Size Adjustment

▪▪▪▪▪▪▪ 11 • The XDS instrument appears just like a network printer (or other peripheral device) on the LAN system. It generates no signals, and onl

Page 36

128 ▪▪▪▪▪▪▪ Vision displays the wavelength regions for test. For reflectance instruments, these regions give a good overall picture of instrument per

Page 37

▪▪▪▪▪▪▪ 129 Insert the 99% standard into the tray as shown, with the label parallel to the long axis of the tray. When the 99% standard has been sca

Page 38

130 ▪▪▪▪▪▪▪ The five standards are plotted out in the USP format, showing repeatable response for each standard. The x-axis, A REF, is from the initi

Page 39

▪▪▪▪▪▪▪ 131 To start Gain Adjust, click on Diagnostics, then Gain Adjust. The instrument is connected and in communication for this to function. Th

Page 40

132 ▪▪▪▪▪▪▪ From the Diagnostics menu bar, select Low Flux Test. Vision asks if the user wishes to use an external sample (standard) for the test.

Page 41

▪▪▪▪▪▪▪ 133 Click on the tab marked “Summary” to see the summarized results as compared to acceptance specifications. Vision reports a pass or fail

Page 42 - 5.6 Universal Tablet Tray

134 ▪▪▪▪▪▪▪ 9 Instrument Maintenance Instrument maintenance is quite simple on the XDS MasterLab Analyzer. The optical enclosure is sealed to preven

Page 43

▪▪▪▪▪▪▪ 135 9.2 Fan Filter Replacement The fan filter should be inspected at least monthly. (If installed in a dusty or dirty environment, it should

Page 44 - 5.7 Tablet Tray Unloader

136 ▪▪▪▪▪▪▪ Using the new filter material, gently insert it into the filter cover. It should fit easily in any direction, as it is symmetrical. Be s

Page 45 - 5.8 Iris Adapter

▪▪▪▪▪▪▪ 137 Turn off power at the power switch located on the back panel. Unplug the power cord. While this is not essential, it makes access easier

Page 46

12 ▪▪▪▪▪▪▪ 3. 3. Upon being powered up on a LAN, the XDS instrument requests a “dynamic” IP address from the DHCP server which controls the LAN. Thi

Page 47

138 ▪▪▪▪▪▪▪ This orientation photo shows the area inside the back cover. The lamp box is at the lower left. The fan and heat sinks, shown to the righ

Page 48

▪▪▪▪▪▪▪ 139 Gently grasp the corners of the lamp mounting plate. Push in about 1mm, then rotate clock-wise (to the right) about 5 mm. The lamp mounti

Page 49 - 6.1 Entry into Vision

140 ▪▪▪▪▪▪▪ Position the lamp with the black arrow up as shown. Gently insert the lamp and mounting plate into the lamp housing as shown. Engage the

Page 50

▪▪▪▪▪▪▪ 141 Carefully re-install the back cover, lining up the captive screws with the mating threaded holes. Tighten gently by hand. Do not over-ti

Page 51

142 ▪▪▪▪▪▪▪ Perform tests as specified in the prior sections of this manual. These normally include the following: • Wavelength Linearization • Per

Page 52

▪▪▪▪▪▪▪ 143 Remove the power cord for access to the fuse. Use a tool (such as an Allen wrench) to pull the housing open. There is a small slot, und

Page 53

144 ▪▪▪▪▪▪▪ When finished, re-install the power cord. Turn on AC power and re-establish instrument communication. 9.5 Maintenance Log Vision provid

Page 54

▪▪▪▪▪▪▪ 145 This screen shows entries, sorted by record number. To re-sort by another parameter, click on the column heading. 9.6 Window Cleaning T

Page 55

146 ▪▪▪▪▪▪▪ After cleaning, allow any residual alcohol to evaporate off. Then use the sample brush from the accessory kit to gently brush away any li

Page 56

▪▪▪▪▪▪▪ 147 10 Validation Tools Validation is an overriding concern in the pharmaceutical marketplace. In the United States, manufacturers must foll

Page 57

▪▪▪▪▪▪▪ 13 3.4.1 Microsoft Windows Firewalls The Microsoft Windows® Firewall on the PC may interfere with Vision communication. To assure communicat

Page 58

148 ▪▪▪▪▪▪▪ Wavelength Accuracy, Photometric Certification, and Wavelength Certification. These factory tests are performed under controlled, ideal c

Page 59 - 6.2.1 Sample ID Entry

▪▪▪▪▪▪▪ 149 Independent assessment, not performed by instrument user • Meets independent test requirements of CFR 21 and most auditing methods. Ful

Page 60

150 ▪▪▪▪▪▪▪ the pharmaceutical industry. Documents were selected (from the enormous quantity of records in our logs) based upon the patterns set duri

Page 61

▪▪▪▪▪▪▪ 151 This screen shows many of the setup items to be set by the System Manager for 21 CFR Part 11 Compliance. To access this screen in Vision,

Page 62

152 ▪▪▪▪▪▪▪ 11 Safety and Electrical Certification The XDS MasterLab Analyzer and all associated components have been tested for CE (Communite Europ

Page 63

▪▪▪▪▪▪▪ 153 12 Troubleshooting The XDS MasterLab Analyzer is a dependable, trouble-free instrument, designed for many years of service in your labor

Page 64 - 6.2.2 Order of Sampling

154 ▪▪▪▪▪▪▪ 2 No communication between Vision and the instrument. 1. Verify that the RJ-45 cable is plugged in at both the instrument and at the net

Page 65

▪▪▪▪▪▪▪ 155 3 MasterLab Analyzer appears to “lock up” upon successful connection. 1. Upon successful connection, the instrument “lost track” of whic

Page 66

156 ▪▪▪▪▪▪▪ 5 Instrument fails Wavelength Linearization 1. Instrument Reference is not in correct position. Verify that reference paddle is visible

Page 67

▪▪▪▪▪▪▪ 157 7 Instrument fails Instrument Calibration. 1. Verify that the Wavelength Standard is properly positioned in the holder when prompted by

Page 68

14 ▪▪▪▪▪▪▪ Internet Protocol (IP) and it is the first version of the protocol to be widely deployed. Together with IPv6, it is at the core of standar

Page 69

158 ▪▪▪▪▪▪▪ 9 Instrument cooling fans are operating at maximum rate. 1. Verify that air intake on side is not blocked by other equipment. Leave at l

Page 70

▪▪▪▪▪▪▪ 159 12 X-Y Sampling Mechanism crashes into Iris Adapter, causing Limit Switch Error Remove Iris Adapter. Click “Use Iris” in Vial Dialog box.

Page 71

160 ▪▪▪▪▪▪▪ 16 Vision reports “Auto Linearization failed” Auto Linearization is a program to maintain correct wavelength position of the instrument.

Page 72

▪▪▪▪▪▪▪ 161 13 Lifting and transporting 1 Always separate the module from the monochromator (shown) before lifting or transporting the instrument. W

Page 73

162 ▪▪▪▪▪▪▪ 13.1 Instructions for MasterLab® Shipping Position When shipping the MasterLab Analyzer, it is vital to set the sample carriage to the “

Page 74

▪▪▪▪▪▪▪ 163 When re-entering the project to run the instrument, be sure to select the correct “Use Tray” or “Use Iris” setting in Vision. This assure

Page 75

164 ▪▪▪▪▪▪▪ 14 Index Accessory Kit 21 Aperture 26, 27, 76 Assembly 21, 46 Auto-Linearization 50, 68, 102, 155, 160 Bias 89, 122 Boat 28, 50, 6

Page 76

▪▪▪▪▪▪▪ 165 Vibration 8 Voltage 160 Wavelength Certification 69, 83, 90, 93, 94, 103, 116, 118, 123, 124, 125, 148 Wavelength Linearization 74, 7

Page 77

▪▪▪▪▪▪▪ 15 2. Highlight “NIRSystems XDS-series Instrument Driver” as shown, then click on “Configure”. Information: At this point, Vision requests a

Page 78

16 ▪▪▪▪▪▪▪ 1. Vision cannot see any instrument on the connection path. Click on the “down arrow” at the right side of the empty field to see if inst

Page 79

▪▪▪▪▪▪▪ 17 • Click on Start, then Control Panel • Double-click on Network Connections • Double-click on Local Area Connection • Click on Local Ar

Page 81

18 ▪▪▪▪▪▪▪ Network Solution 2: Returning to the Local Area Connection Status dialog box, note these items for the computer: • Address Type: (should

Page 82

▪▪▪▪▪▪▪ 19 Network Solution 4: Verify that the Firewall on the computer has Vision loaded as an “exception”. This is located under “Control Panel”,

Page 83

20 ▪▪▪▪▪▪▪ 3.6.2 Direct Connection Troubleshooting Overview If no XDS instrument shows as “available”, the computer may need to be configured for t

Page 84

▪▪▪▪▪▪▪ 21 4 Assembly of the Instrument The XDS Analyzer will be assembled and installed by a trained representative of Metrohm. This person will pe

Page 85

22 ▪▪▪▪▪▪▪ 1 Load Vision Spectral Analysis Software onto the computer designated to operate the XDS instrument. 2 Place the monochromator on the lab

Page 86

▪▪▪▪▪▪▪ 23 5 Insert the AC power cable into the AC power block as shown. 6 Attach the RJ-45 cable to the network connector on the instrument. If usi

Page 87

24 ▪▪▪▪▪▪▪ 9 Lift the release handle on the monochromator and engage the module “catches” to the locking togs on the monochromator. (Module not shown

Page 88

▪▪▪▪▪▪▪ 25 13 When all the above assembly is finished, turn on the power switch on the monochromator. It is located on the lower surface, on the righ

Page 89 - 7.2.1 Performance Test

26 ▪▪▪▪▪▪▪ 5 MasterLab Analyzer Overview The XDS MasterLab Analyzer is designed for several types of versatile analysis, from tablet sampling, to vi

Page 90

▪▪▪▪▪▪▪ 27 Large Sample Cell, Reflectance • Detector Selection: Reflectance • Wavelength Range: 400-2500 nm • Preferred Spot Size: 17.25 mm • Ape

Page 91

Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 [email protected] www.metrohm.com NIRS XDS MasterLab Analyz

Page 92

28 ▪▪▪▪▪▪▪ Transmission Reflectance Use Tray Use for multiple tablets, scanning in transmission. The default spot size (9.5mm) is appropriate for

Page 93

▪▪▪▪▪▪▪ 29 5.1 Introduction to Tablet Analysis The MasterLab Analyzer is designed to provide sampling for trays of tablets, like that shown at right

Page 94

30 ▪▪▪▪▪▪▪ that in tablet transmission, some sample positions are not available because the reference standard takes up one or two tablet positions.

Page 95

▪▪▪▪▪▪▪ 31 5.2 Creation of a Custom Vial Size Most users will work with one of the standard vial sizes (or tablet layouts) offered. However, for set

Page 96 - 7.2.3 Photometric Test

32 ▪▪▪▪▪▪▪ A “Tray ID” must be assigned. The normal convention is to use the diameter, then the number of vials in a row, and a column. You may use y

Page 97

▪▪▪▪▪▪▪ 33 These dimensions are the center of each vial opening in the tray. Vision uses these coordinates to position each vial during sampling. Not

Page 98

34 ▪▪▪▪▪▪▪ The adjustment ranges from 9.5mm to 17.25mm, in adjustment steps of 0.25mm. The effective scanned sample areas at various spot sizes are

Page 99

▪▪▪▪▪▪▪ 35 When using the default spot size of 17.25mm, the sample diameter should be at least 21.6mm (0.85 inch) as represented by the gray circle i

Page 100

36 ▪▪▪▪▪▪▪ To insert the sample tray into the instrument, note the round guides on the sample carrier. They are shown in this photo – the pencil poin

Page 101 - 7.2.4 Gain Test

▪▪▪▪▪▪▪ 37 1. Remove the black covers from the adjustment post on the instrument, and the adjustment screw packed with the standards tray. 2. Thr

Page 102 - 7.2.5 Low Flux Test

Teachware Metrohm AG CH-9100 Herisau [email protected] This documentation is protected by copyright. All rights reserved. Although all

Page 103

38 ▪▪▪▪▪▪▪ 5. The vertical post of the adjustment is marked in 1mm (about 0.040”) increments. The rotary knob is marked in 50 increments. One markin

Page 104

▪▪▪▪▪▪▪ 39 Once set, the height will remain constant, even when the transmission detector is moved to the storage position as shown. When rotated bac

Page 105

40 ▪▪▪▪▪▪▪ 5.6 Universal Tablet Tray Part Number 3013-243-01 The XDS MasterLab® Analyzer is shipped with one “Universal Tablet Tray” which may be us

Page 106

▪▪▪▪▪▪▪ 41 of the Universal Tablet Tray. (Remember that spectral intensity changes somewhat with detector height.) Bore sizes (diameter), for tablet

Page 107

42 ▪▪▪▪▪▪▪ 5.7 Tablet Tray Unloader Because tablets fit snugly in the tray, and may be difficult to remove, a device has been designed to make unloa

Page 108

▪▪▪▪▪▪▪ 43 The vacuum lifting device is easily controlled by blocking or releasing the hole on the wand. Cover the hole to pick up the tablet. To rel

Page 109

44 ▪▪▪▪▪▪▪ When using the Iris Adapter, Vision provides a command to move the sample carriage to a neutral position, as shown at right. First, remove

Page 110

▪▪▪▪▪▪▪ 45 When ready, and all interfering items are removed, click “OK” in this dialog box. When scanning stationary objects, the “Use Iris” select

Page 111

46 ▪▪▪▪▪▪▪ Use the iris to center samples in the window area. Push the handle to the right to center the sample, then to the left to move the iris fr

Page 112

▪▪▪▪▪▪▪ 47 6 Vision Software: Connection to the Instrument This section describes communication between the computer (with Vision Software loaded) a

Page 113

▪▪▪▪▪▪▪ 3 Table of contents 1 Introduction to XDS™ MasterLab Analyzer ... 5 2 Site

Page 114

48 ▪▪▪▪▪▪▪ A new project must be created. The “project” is used to store data and calibrations for a given type of analysis. Multiple projects may b

Page 115

▪▪▪▪▪▪▪ 49 This box directs the user to select the instrument IP (Internet Protocol) address. Use the drop-down arrow of the IP Address box, and sele

Page 116

50 ▪▪▪▪▪▪▪ Verify that these selections are clicked: Instrument must stabilize before data acquisition Use Auto-Linearization Reference Standardizati

Page 117

▪▪▪▪▪▪▪ 51 If the user clicks on “Modify/ View” from the Select Tray Dialog box, this screen is shown. If “New” is clicked, all fields are blank, for

Page 118

52 ▪▪▪▪▪▪▪ The instrument “self-identifies” as an XDS system with the Tablet Module. The user should enter a logical name for the DCM. (Spaces are OK

Page 119

▪▪▪▪▪▪▪ 53 Next, select this DCM from the selection box. Click “OK” to connect to the instrument. Once connected, verify that the amber “Communicatio

Page 120

54 ▪▪▪▪▪▪▪ If a Reference Standardization does not exist for the selected configuration, Vision will display this message. The user should perform b

Page 121

▪▪▪▪▪▪▪ 55 instrument operation. Please proceed to the next step. In tablet transmission, Vision stores yet another Reference Standardization, for th

Page 122 - 8.2.1 Performance Test

56 ▪▪▪▪▪▪▪ Note that the Vial Dialog box is also called when either of these actions is performed: 1. The user clicks on the “Sample” Icon, or 2.

Page 123

▪▪▪▪▪▪▪ 57 Also note that 20 vials are shown, in a 4 x 5 configuration. This came from the Select Tray Dialog box. (It is shown at right, as a remin

Page 124

4 ▪▪▪▪▪▪▪ 7.1.2 Reference Standardization, Tablet Transmission ... 76 7.1.3 Instrument Calibration

Page 125

58 ▪▪▪▪▪▪▪ Entry Method 2: Double-click on the upper part of the vial position, which brings up the box shown: The Sample ID may be typed into this

Page 126

▪▪▪▪▪▪▪ 59 The cell entry method is shown above, highlighted by the red border. The user may type a Sample ID, then press “Enter” to move down to th

Page 127

60 ▪▪▪▪▪▪▪ Highlight the Sample ID fields to be pasted as shown above. Click on the “Paste” button in the dialog box. (It is outlined in red below.)

Page 128

▪▪▪▪▪▪▪ 61 There is a field for comments beside each sample. To see this field, click the small arrow at the bottom right, and the field will be visi

Page 129

62 ▪▪▪▪▪▪▪ To delete multiple Sample ID names, highlight those fields, and use the “Delete” key on the PC keyboard. Vision will ask for confirmation,

Page 130

▪▪▪▪▪▪▪ 63 below, this is automatic. At the bottom of the Vial Dialog screen is a checkbox marked “Use Order”. If this box is checked, the MasterLa

Page 131

64 ▪▪▪▪▪▪▪ In the example above, the order is out of sequence for samples, 7, 9, and 16. If “Use Order” is left unchecked, Vision proceeds in order

Page 132 - 8.2.4 Gain Test

▪▪▪▪▪▪▪ 65 This setting defines the number of sample spectra to be taken per reference scan. With multiple samples, it is cumbersome to take a refere

Page 133 - 8.2.5 Low Flux Test

66 ▪▪▪▪▪▪▪ Use Iris: If the user needs to scan a sample using the Iris Adapter, this box should be clicked. (This resets the choice made in the “Sel

Page 134

▪▪▪▪▪▪▪ 67 If the user needs to scan a sample using the large Sample Cell (boat), this box should be clicked. (This resets the choice made in the “Se

Page 135

▪▪▪▪▪▪▪ 5 1 Introduction to XDS™ MasterLab Analyzer Thank you for selecting the XDS MasterLab Analyzer, manufactured by FOSS. This instrument is the

Page 136 - 9 Instrument Maintenance

68 ▪▪▪▪▪▪▪ 7 Instrument Diagnostics, Tablet Transmission Vision provides diagnostics for instrument setup. These diagnostics must be performed befor

Page 137 - 9.2 Fan Filter Replacement

▪▪▪▪▪▪▪ 69 This is a method to adjust the instrument wavelength profile to an external, traceable wavelength standard. This feature must be used to a

Page 138 - 9.3 Lamp Replacement

70 ▪▪▪▪▪▪▪ A new DCM screen will be shown. The “Method” name will be blank, and the “Detector” may be marked as “Reflectance”. Proceed with these st

Page 139

▪▪▪▪▪▪▪ 71 To verify this connection, the user may click on Diagnostics, Instrument Configuration. Note that the Amplifier Type is listed as “Transm

Page 140

72 ▪▪▪▪▪▪▪ It is easier to install the aperture when the sample carriage is in the “Iris Position”. The unit may be in this position from previous w

Page 141

▪▪▪▪▪▪▪ 73 To re-install the tablet transmission reference, mount it onto the rail as shown, and screw down the captive thumbscrew about 4-5 turns.

Page 142

74 ▪▪▪▪▪▪▪ The directions that follow include spectroscopic calibrations, which are used to “normalize” the instrument, to minimize differences betwe

Page 143

▪▪▪▪▪▪▪ 75 These peak positions are not meant to be traceable, as the true wavelength calibration of the instrument is done on an external standard,

Page 144 - 9.4 Fuse Replacement

76 ▪▪▪▪▪▪▪ If Wavelength Linearization fails for any reason, check these items: • Tablet Transmission Detector in correct position • Tablet Transmi

Page 145

▪▪▪▪▪▪▪ 77 When Vision first communicates with the instrument (or after disconnection and re-connection) this screen is shown. Vision must establish

Page 146 - 9.5 Maintenance Log

6 ▪▪▪▪▪▪▪ the sections on Reference Standardization. This instrument uses near-infrared (NIR) spectral energy to illuminate the sample. By measuring

Page 147 - 9.6 Window Cleaning

78 ▪▪▪▪▪▪▪ Vision plots the measured difference between the instrument tablet transmission reference and the certified standard in the Transmission S

Page 148

▪▪▪▪▪▪▪ 79 Vision plots the spectrum of the final corrected reference spectrum in dark green. This spectrum should appear exactly the same as the cal

Page 149 - 10 Validation Tools

80 ▪▪▪▪▪▪▪ From the menu bar, click on Diagnostics, then Instrument Calibration. Vision requests that the aperture and Tablet Transmission Standard

Page 150

▪▪▪▪▪▪▪ 81 Vision requests the “Standard File” for the Certified Wavelength Standard. Place the mini-CD into the CD drive, and select the standard f

Page 151

82 ▪▪▪▪▪▪▪ At the end of the test this dialog box is displayed. Print confirmation of test results if needed for a log book or other file. Click “Pri

Page 152

▪▪▪▪▪▪▪ 83 to illustrate that the wavelength peaks of SRM-2035 remain constant, even when neutral density filters are added to increase absorbance le

Page 153

84 ▪▪▪▪▪▪▪ under Configure, Options. That is, if the IPV Setup file is acquired with Reference Standardization switched on, and with Window Correctio

Page 154

▪▪▪▪▪▪▪ 85 Vision verifies the tray type – select the Tablet Transmission Standard, as highlighted. Verify “Use Tray”. Click on “Select”, then “Close

Page 155 - 12 Troubleshooting

86 ▪▪▪▪▪▪▪ Vision takes an instrument reference. The reference spectrum is shown in red. Note that Reference Standardization is not shown as applied

Page 156

▪▪▪▪▪▪▪ 87 7.2 Evaluation Diagnostics, Tablet Transmission Evaluation Diagnostics are used to verify that the instrument is operating within allowab

Page 157

▪▪▪▪▪▪▪ 7 The instrument enclosure is completely sealed to prevent contamination by dust or other substances. The cooling fans operate outside the ma

Page 158

88 ▪▪▪▪▪▪▪ Because only the right two quadrants of the screen will be updated, many users “pull the screen over” to see better, as shown on the next

Page 159

▪▪▪▪▪▪▪ 89 When Performance Test is finished, a message box is displayed to indicate test completion and status. All measured values are compared wit

Page 160

90 ▪▪▪▪▪▪▪ protected and are moved by software command, transparent to the user. When Performance Test is run, the relative wavelength positions and

Page 161

▪▪▪▪▪▪▪ 91 Vision asks the user to select the Standard File. This file is on the mini-CD, as shown. This file may be copied into the C:\Vision direc

Page 162

92 ▪▪▪▪▪▪▪ Double-click the lower right quadrant to see the full report. The first tab shows the Wavelength position, as found. Only one is used w

Page 163 - 13 Lifting and transporting

▪▪▪▪▪▪▪ 93 Note that the peak position is shown only to two digits after the decimal point. Any further digits are not considered significant, and ar

Page 164

94 ▪▪▪▪▪▪▪ In some cases the user may see the message shown. This indicates that the instrument has passed Wavelength Certification on the NIST-defin

Page 165

▪▪▪▪▪▪▪ 95 Vision requests a “Standard File”. For Photometric Test, use the TSSVxxxx.da file stored in the Vision directory. This file was created du

Page 166 - 14 Index

96 ▪▪▪▪▪▪▪ Vision displays the wavelength regions for test. For tablet transmission, these regions give a good overall picture of instrument performa

Page 167

▪▪▪▪▪▪▪ 97 Vision verifies the tray layout as shown. Click on “OK”. Vision acquires a reference scan. Next, the T02 is scanned. It is reported at ri

Comments to this Manuals

No comments