1.3 Titration modes – Measuring modes – Dosing commands
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
904 Titrando
MEAS
The following measuring modes can be selected for measurements:
pH (pH measurement)
U (potentiometric voltage measurement)
Ipol (voltametric measurement with selectable polarization cur-
rent)
Upol (amperometric measurement with selectable polarization
voltage)
Conc (concentration measurement with or without standard
addition)
T (temperature measurement)
STDADD
The measuring modes for standard addition are listed separately only in
tiamo™. In Touch Control they are integrated in the measuring mode
MEAS Conc.
The following measuring modes can be selected for measurements:
auto (automatic addition of the standard addition solution by
specifying a potential difference)
dos (automatic addition of the standard addition solution by
specifying the individual volume increments)
man (manual addition of the standard addition solution)
CAL
Electrode calibration.
Measuring mode:
pH (calibration of pH electrodes)
Conc (calibration of ion-selective electrodes)
ELT
Electrode test for pH electrodes.
This mode is listed separately only in tiamo™. In Touch Control, the
electrode test is a component part of the CAL calibration mode.
Dosing commands
The following commands for dosing can be selected:
PREP (rinsing the cylinder and tubings of an exchange unit or
dosing unit)
EMPTY (emptying the cylinder and tubings of a dosing unit)
ADD (dosing a specified volume)
LQH (carrying out complex dosing tasks with a Dosino)