1 Introduction
904 Titrando
■■■■■■■■
3
1.3 Titration modes – Measuring modes – Dosing com-
mands
The 904 Titrando supports the following titration modes, measuring
modes and dosing commands:
DET
Dynamic equivalence point titration. The reagent addition is carried out
in variable volume steps.
Measuring modes:
pH (pH measurement)
U (potentiometric voltage measurement)
Ipol (voltametric measurement with selectable polarization cur-
rent)
Upol (amperometric measurement with selectable polarization
voltage)
MET
Monotonic equivalence point titration. The reagent addition is carried
out in constant volume steps.
Measuring modes:
pH (pH measurement)
U (potentiometric voltage measurement)
Ipol (voltametric measurement with selectable polarization cur-
rent)
Upol (amperometric measurement with selectable polarization
voltage)
SET
Endpoint titration at one or two specified endpoints.
Measuring modes:
pH (pH measurement)
U (potentiometric voltage measurement)
Ipol (voltametric measurement with selectable polarization cur-
rent)
Upol (amperometric measurement with selectable polarization
voltage)