Page view 90
6.4 819 IC Detector
IC Net 2.3
81
Remote configuration
On the Remote configuration subtab user-defined remote com-
mands can be defined, which can be inserted into a time pro-
gram after being activated with <Activate>.
Name (1
st
column) User-definable name of the remote com-
mand (e.g. Start_766).
Remote command (2
nd
column)
Setting the remote output lines 18.
0 (line off, inactive, open)
1 (line on, active, 0 V)
* (leave line in current status)
For entry of the first value, enter 1, 0 or *.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter 1, 0 or *.
<Add> Add new remote command.
<Delete> Delete selected remote command.
<Activate> Activate the defined remote commands for
insertion into the time program.
Links
The Links tab of the 819 IC Detector is used for COM port selection
and settings (details see section 5.3.4 Links).