5.3 830 IC Interface
IC Net 2.3
51
Label Optional label to name the interface with
maximum 8 characters.
ROM version Instrument program version number.
Measuring per second Number of data points measured per s.
Entry range: 10, 20, 30, 50, 60 points/s
Range Range for AD converter.
Example: converts ± 2500 mV to 2
24
bits
Entry range: 2500, 1250, 625, 312.5, 156.25,
78.125, 39.062 mV
<SET> Send current parameters to 830 IC Inter-
face.
RS settings
The RS settings tab of the 830 IC Interface window contains RS232
parameters for the 830 IC Interface.