4 Short operating tutorial
826/827 pH meter, Instructions for Use 17
4 Short operating tutorial
In this section the necessary steps for carrying out a simple pH and ion
measurement with calibration are described. The instructions are li-
mited to those steps that are absolutely necessary and will enable you
to carry out your first measurements with the pH meter directly. The op-
erating principles are described in Section 3.3.
4.1 Requirements
The following instruments, accessories and solutions are required for
carrying out the pH calibration and measurement described below:
826 pH mobile (2.826.0XX0) or 827 pH lab (2.827.011X,
2.827.021X)
pH electrode
Calibration buffers
Metrohm buffer solutions pH 4.00 and pH 7.00
The calibration parameters for the pH mode are set for a calibration us-
ing two Metrohm buffers as default (see Section 6.1.2). If you want to
use other buffers then the corresponding buffer type must be entered.
4.2 pH calibration
1 Start calibration with first buffer
Start calibration with <CAL>
Immerse pH electrode in buffer solution pH 7
and confirm with
<OK>
If a temperature sensor is connected, the calibration tempera-
ture is being determined
If no temperature sensor is connected:
enter the temperature with the arrow keys and confirm with
<OK>
The first buffer is being measured
The following message is displayed after successful mea-
surement:
change buffer <OK>
2 Continue calibration with second buffer
Remove pH electrode from first buffer and rinse with water
Immerse pH electrode in second buffer solution pH 4 and
continue calibration sequence with <OK>
The second buffer is being measured