Metrohm 930 Compact IC Flex Oven/ChS/PP/Deg User Manual Page 62

  • Download
  • Add to my manuals
  • Print
  • Page
    / 118
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 61
3.20 Connecting the separation column
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
930 Compact IC Flex Oven/ChS/PP/Deg (2.930.2360)
2
Installing the input of the separation column
There are 3 possibilities:
Attaching the bottom end of the separation column directly to
the guard column.
or
If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or
If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).
3
Rinsing the separation column
Place a beaker under the separation column's outlet.
Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leaflet
On
Rinse the separation column with eluent for approximately 10
minutes.
Use manual control in MagIC Net to stop the high-pressure pump
again: Off.
4
Removing the coupling
Remove the coupling (6.2744.040) from the column output capil-
lary.
5
Installing the output of the separation column
Fasten the column output capillary to the upper end of the sepa-
ration column using a PEEK pressure screw (6.2744.070).
6
Inserting the separation column
Insert the separation column with chip into the column holder
until you hear it snap in place.
The separation column is now detected by MagIC Net.