5.4 Performing a sample series
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
885 Compact Oven SC
5.4 Performing a sample series
Blanks should always be at the first sample position so that the blank val-
ues determined are available with every sample processing.
5.4.1 Starting the sample series
Starting a sample series
A suitable method must be loaded before a sample series is started (see
Chapter 5.3.3, page 32). The necessary parameters ) can then be modi-
fied.
1
Define the sample series
Press the [START] key.
You can now enter the quantity and the first rack position of the
samples to be processed.
2
Enter the number of samples
■ Select Number of samples and press [OK].
■ Enter the number of samples including the blanks.
■ Close the input dialog with [BACK] or Accept.
Note
Take care to ensure that the number of samples matches the num-
ber of sample data entered in the titrator.
3
Enter the rack position of the first sample
■ Select Next sample pos. and press [OK].
■ Enter the starting position of the sample series.
■ Close the input dialog with [BACK] or Accept.
The value for the number of samples remains saved for the next sam-
ple series. The position of the first sample is increased with each
method run.
You can still cancel the start of the sample series at this time with
[BACK] or [STOP].
4
Close the sample series dialog
Close the dialog with the [BACK] key.