Summary of Contents

Page 1 - 753 Suppressor Module

CH-9101 Herisau/SwitzerlandTel. ++41 71 353 85 85Fax ++41 71 353 89 01Internet www.metrohm.chE-Mail [email protected] Suppressor ModuleSuppressor Mo

Page 2 - Table of contents

2.1 Setting up the instrument753 Suppressor Module / 8.753.1003 Instructions for Use72 Installation2.1 Setting up the instrument2.1.1 PackagingThe

Page 3 - List of figures

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use82.2 Mains connectionFollow the instructions below for connecting to the power

Page 4 - 1 Introduction

2.2 Mains connection753 Suppressor Module / 8.753.1003 Instructions for Use91313220 -- 240 V 100 -- 120 V 19192020220 -- 240 V 100 -- 120 V Fig. 4:

Page 5 - 1.2 Parts and controls

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use102.3 Connection to the 690 Ion Chromatograph2.3.1 Electrical connectionIn ord

Page 6 - 2020 161617171818

2.3 Connection to the 690 Ion Chromatograph753 Suppressor Module / 8.753.1003 Instructions for Use112Connect suppressor block• Remove the plastic st

Page 7 - 2121 2222 2323

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use127Attach pump tubings• Loosen both tubing cartridges 44 from the holding clamp

Page 8 - >PARAM/detector

2.3 Connection to the 690 Ion Chromatograph753 Suppressor Module / 8.753.1003 Instructions for Use133030 3232 3333 3434 44 55 3434 3232 25/2825/28101

Page 9 - 1.4 Safety notes

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use14• Pull outlet capillary 2727 marked with "Waste" at suppressorconne

Page 10 - 2 Installation

2.4 Connection to the 732/733 IC system753 Suppressor Module / 8.753.1003 Instructions for Use152.4 Connection to the 732/733 IC system2.4.1 Electr

Page 11 - 2.2 Mains connection

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use161Insert suppressor block• 733.0010 IC Separation Center (1-channel system):Pl

Page 12 - 100 -- 120 V

Table of contents753 Suppressor Module / 8.753.1003 Instructions for UseITable of contents1 Introduction...

Page 13 - Cable of suppressor block

2.4 Connection to the 732/733 IC system753 Suppressor Module / 8.753.1003 Instructions for Use178Fix connection suppressor – detector block• Insert

Page 14 - 29292525

2 Installation753 Suppressor Module / 8.753.1003 Instructions for Use1811Suppressor connection 3: H2O• Pull the inlet capillary 2828 marked with “H

Page 15

3.1 Handling the suppressor753 Suppressor Module / 8.753.1003 Instructions for Use193 Operation3.1 Handling the suppressorGeneralThe Metrohm Suppre

Page 16

3 Operation753 Suppressor Module / 8.753.1003 Instructions for Use203.2 Manual operationThe 753 Suppressor Module cannot be manually operated unless

Page 17

3.3 Operation via 690 Ion Chromatograph753 Suppressor Module / 8.753.1003 Instructions for Use213.3 Operation via 690 Ion ChromatographThe 753 Suppr

Page 18 - Cable of

3 Operation753 Suppressor Module / 8.753.1003 Instructions for Use223.4 Operation via 732 IC DetectorThe 753 Suppressor Module cannot be remotely op

Page 19

3.4 Operation via 732 IC Detector753 Suppressor Module / 8.753.1003 Instructions for Use23Switch pump on/off in a programIn the 732 IC Detector the &

Page 20

3 Operation753 Suppressor Module / 8.753.1003 Instructions for Use24Rotate:>PROGRAM/edit XXX.X minremote T1TTTTTTResetoutput lead:>PR

Page 21

4.1 Malfunctions and their rectification753 Suppressor Module / 8.753.1003 Instructions for Use254 Malfunctions – Maintenance4.1 Malfunctions and t

Page 22 - 3 Operation

4 Malfunctions – Maintenance753 Suppressor Module / 8.753.1003 Instructions for Use26Suppressor moves tonext position but displaylamp 77 no longer li

Page 23 - 3.2 Manual operation

Table of contents753 Suppressor Module / 8.753.1003 Instructions for UseII5 Appendix ...

Page 24 - Rotate suppressor

4.2 Maintenance and servicing753 Suppressor Module / 8.753.1003 Instructions for Use274.2 Maintenance and servicing4.2.1 CareThe 753 Suppressor Mod

Page 25 - >SETUP/output

4 Malfunctions – Maintenance753 Suppressor Module / 8.753.1003 Instructions for Use284.2.4 Exchanging the pump tubingPump tubings are consumable mat

Page 26 - 0TTTTTTT

4.2 Maintenance and servicing753 Suppressor Module / 8.753.1003 Instructions for Use294.2.5 Regeneration of the suppressorIf the suppressor units ar

Page 27 - Suppressor position display

4 Malfunctions – Maintenance753 Suppressor Module / 8.753.1003 Instructions for Use304.2.6 Cleaning the suppressorCleaning the suppressor may be nec

Page 28 - 4 Malfunctions – Maintenance

4.2 Maintenance and servicing753 Suppressor Module / 8.753.1003 Instructions for Use313838373736363535 Fig. 9: Assembling the suppressor3535 Screw n

Page 29

4 Malfunctions – Maintenance753 Suppressor Module / 8.753.1003 Instructions for Use324.2.7 Exchanging the suppressorExchanging the suppressor in the

Page 30 - 4.2.3 Shutdown

4.2 Maintenance and servicing753 Suppressor Module / 8.753.1003 Instructions for Use336Insert connection piece• Insert new connection piece 3636 in

Page 31

4 Malfunctions – Maintenance753 Suppressor Module / 8.753.1003 Instructions for Use34• By rotating the suppressor block or with the aid of a mirrorc

Page 32

5.1 Technical data753 Suppressor Module / 8.753.1003 Instructions for Use355 Appendix5.1 Technical dataPumpPump type 2-channel peristaltic pump wit

Page 33

5 Appendix753 Suppressor Module / 8.753.1003 Instructions for Use36Mains connectionMains voltage 115 V: 100...120 V ± 10 %230 V: 220...240 V ± 10 %

Page 34 - 3838373736363535

1.1 Instrument description753 Suppressor Module / 8.753.1003 Instructions for Use11 Introduction1.1 Instrument descriptionThe 753 Suppressor Module

Page 35

5.2 Standard equipment753 Suppressor Module / 8.753.1003 Instructions for Use375.2 Standard equipmentSubject to changes!All dimensions are given in

Page 36

5 Appendix753 Suppressor Module / 8.753.1003 Instructions for Use385.3 Optional accessoriesOrder No. Description6.2143.210 Connecting cableConnectin

Page 37

5.4 Warranty and conformity753 Suppressor Module / 8.753.1003 Instructions for Use395.4 Warranty and conformity5.4.1 WarrantyThe warranty on our pr

Page 38 - 5 Appendix

5 Appendix753 Suppressor Module / 8.753.1003 Instructions for Use405.4.2 EU Declaration of conformityEU Declaration of ConformityThe METROHM AG comp

Page 39

5.4 Warranty and conformity753 Suppressor Module / 8.753.1003 Instructions for Use415.4.3 Certificate of conformity and system validationCertificate

Page 40 - 5.2 Standard equipment

5 Appendix753 Suppressor Module / 8.753.1003 Instructions for Use42

Page 41 - 5.3 Optional accessories

5.5 Index753 Suppressor Module / 8.753.1003 Instructions for Use435.5 IndexAAdjustment screw 1818Braking range adjustment...33Figure...

Page 42 - 5.4 Warranty and conformity

5 Appendix753 Suppressor Module / 8.753.1003 Instructions for Use44Operation via690 Ion Chromatograph...21Operation via 732 IC Detecto

Page 43 - EU Declaration of Conformity

1 Introduction753 Suppressor Module / 8.753.1003 Instructions for Use21.2 Parts and controlsSuppressor Module753 MetrohmMetrohmON ONOFFOFFREMOTE PUM

Page 44

1.2 Parts and controls753 Suppressor Module / 8.753.1003 Instructions for Use3WARNING - Fire Hazard -with the same type and rating of fuseFor continu

Page 45

1 Introduction753 Suppressor Module / 8.753.1003 Instructions for Use41322121 2222 2323 Fig. 3: Suppressor block (1.753.0100)1.3 Information about

Page 46 - 5.5 Index

1.3 Information about the Instructions for Use753 Suppressor Module / 8.753.1003 Instructions for Use51.3.2 Notation and pictogramsThe following not

Page 47

1 Introduction753 Suppressor Module / 8.753.1003 Instructions for Use61.4 Safety notes1.4.1 Electrical safetyWhile electrical safety in the handlin

Comments to this Manuals