Application Bulletin 321 e
Installation Instructions of TitrIC 7
The TitrIC 7 System is used for the fully automatic analysis of water samples using direct measurement,
titration and ultrafiltration for the ion chromatography measurements. The following measured
parameters are determined in a very short time: temperature, conductivity, pH, acid capacity and, in
parallel, the concentrations of the individual anions and cations. Further Metrohm instruments can be
incorporated in the existing system at any time and used to determine additional parameters.
Analytical sequence
1. Conductivity is measured in the sample vessel on the sample processor.
2. Ultrafiltration inclusivley sample transfer to injection loops of the ion chromatograph.
3. The IC determinations are started automatically.
4. In parallel to the Ultrafiltration the sample is transfered to the external cell of the 855 Robotic
Titrosampler.
5. Temperature, pH and acid capacity (p and m values) are determined.
The whole procedure is controlled by the TitrIC software: The user enters the sample position and
sample identification, MagIC Net and tiamo
TM
method, after which TitrIC transfers all the relevant
information to tiamo
TM
and MagIC Net. When the measurement has been concluded, TitrIC
automatically generates a joint report containing all results. This report can be exported as an Excel file.
Page view 0
1 2 3 4 5 6 ... 15 16

Summary of Contents

Page 1 - Application Bulletin 321 e

Application Bulletin 321 e Installation Instructions of TitrIC 7 The TitrIC 7 System is used for the fully automatic analysis of water samples usin

Page 2 - Table of contents

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 10/16  Conductivity cell (6.0912.110): The measuring cell cable is led through

Page 3 - 1 Delivery package

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 11/16 Confirm the entries twice with <OK> and then mark 855_1 under Device

Page 4 - - optional accessories:

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 12/16 6.2.6 Electrodes  pH electrode: Under Configuration/Sensors click on Edi

Page 5 - 2 Installation

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 13/16 6.2.8 Configuration of the export template Export template: tiamoTM must t

Page 6 - 3 Cable connections

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 14/16 The entry Start counter is obtained at the left by clicking on Determinati

Page 7 - 4 Tubing connections

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 15/16 Input window for the export template The entry Start counter is obtained

Page 8 - 6.1805.030

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 16/16 If TitrIC is shut down then tiamoTM and MagIC Net will shut down automatic

Page 9 - 5 Miscellaneous

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 2/16 Table of contents 1 Delivery package...

Page 10 - 6 Software

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 3/16 1 Delivery package - delivered with TitrIC 7 package: Number Article no.

Page 11 - 6.2.5 Solutions

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 4/16 - optional accessories: 1 6.2041.xxx Sample rack made of PVC 1 6.1432.x

Page 12 - 6.2.7 Methods

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 5/16 2 Installation The following is a detailed description of the TitrIC 7 Sys

Page 13 - Page 13/16

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 6/16 3 Cable connections First all the power cables are connected to the instrum

Page 14 - Æ Shut down tiamo

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 7/16 4 Tubing connections 4.1 External cell The numbers given here (,…) and

Page 15 - 6.4 Configuration of TitrIC

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 8/16 After the screw nipple (6.2730.070) has been screwed into opening 9, the

Page 16 - Page 16/16

Application Bulletin 321 e Installation Instructions of TitrIC 7 Page 9/16 4.3 Overview table of all the liquid handling tubing connections The thr