Application Bulletin 403/1 e
Installation instruction for MVA-21:
894 Professional CVS fully automated for CVS determinations
The «MVA-21» is a fully automated system for the
determination of suppressor and brightener in plating
solutions.
PC controlled operation.
Manual operations:
o Filling of the sample vials on the Sample
Processor with suppressor standard solution,
plating bath samples and rinsing solution.
Automatic addition of 4 solutions: VMS, brightener
concentrate, suppressor concentrate and suppressor
standard or plating bath sample.
Automatic rinsing of the measuring vessel.
Method change during the determination series is
possible.
Applications:
o Analysis of suppressor concentration in plating
baths with CVS.
o Analysis of brightener concentration in plating
baths with CVS.
Fig 1 MVA-21
Content
1. Instruments and accessories .......................................... 2
2. System setup .................................................................. 2
2.1. Electrical connections ......................................... 2
2.2. Tubing connections for rinsing and draining of the
measuring vessel ................................................ 3
2.3. Tubing connections for automatic dosing of
standard solutions and auxiliary solutions ........... 4
3. viva Configuration ........................................................... 5
3.1. Devices 894 Profesional CVS .......................... 5
3.2. Sensors/Electrodes ............................................. 5
3.3. Dosing units ........................................................ 5
3.4. Solutions ............................................................. 6
3.5. Devices 858 Professional Sample Processor for
CVS ..................................................................... 6
3.6. Needle adjustment .............................................. 7
4. viva «Method» ................................................................ 8
4.1. Method run .......................................................... 8
4.2. Evaluation ........................................................... 9
5. Operation...................................................................... 10
5.1. Suppressor determination series ...................... 10
5.2. Brightener determination series ........................ 11
6. Remarks ....................................................................... 12
6.1. Single additive setup ......................................... 12
6.2. Combination of applications .............................. 12
6.3. Storing dosing units........................................... 12
6.4. Storing electrodes ............................................. 12
6.5. Waste container ................................................ 12
Page 1 of 12
Page view 0
1 2 3 4 5 6 ... 11 12

Summary of Contents

Page 1 - Content

Application Bulletin 403/1 e Installation instruction for MVA-21: 894 Professional CVS fully automated for CVS determinations The «MVA-21» is a ful

Page 2 - 2. System setup

Application Bulletin 403/1 e Installation instruction for MVA-21 4.2.4. Results In the «Results» part on the tab «Results» the substance is select

Page 3

Application Bulletin 403/1 e Installation instruction for MVA-21 5.2. Brightener determination series 5.2.1. Manual operation • Dosing Unit «89

Page 4

Application Bulletin 403/1 e Installation instruction for MVA-21 6. Remarks 6.1. Single additive setup If the system is only used for suppressor o

Page 5 - 3. viva Configuration

Application Bulletin 403/1 e Installation instruction for MVA-21 1. Instruments and accessories Quantity Article number 1 2.894.1210 894 Prof

Page 6 - TANDARD SOLUTION

Application Bulletin 403/1 e Installation instruction for MVA-21 2.2. Tubing connections for rinsing and draining of the measuring vessel Fig 3:

Page 7

Application Bulletin 403/1 e Installation instruction for MVA-21 2.3. Tubing connections for automatic dosing of standard solutions and auxiliary

Page 8 - 4. viva «Method»

Application Bulletin 403/1 e Installation instruction for MVA-21 Fig 8: Detailed view for the tubing connections of the dosing unit waste port 3.

Page 9 - – GO TO RINSING POSITION

Application Bulletin 403/1 e Installation instruction for MVA-21 Dosing rate port 1 Maximum mL/L Dosing rate port 2 Maximum mL/L Dosing rate Fi

Page 10 - 5. Operation

Application Bulletin 403/1 e Installation instruction for MVA-21 Swing head Swing position 0 mm Rinse position 0 mm External position Not used

Page 11 - AMPLE has to be selected

Application Bulletin 403/1 e Installation instruction for MVA-21 9. Once the needle is properly adjusted select the tab «Assign position». 10. Th

Page 12 - 6. Remarks

Application Bulletin 403/1 e Installation instruction for MVA-21 4.1.5. Rinsing measuring vessel After each sample the measuring vessel need to be